s'authentifier
version française rss feed
HAL : insu-00713235, version 1

Fiche concise  Récupérer au format
Feedback system of a liquid nitrogen cooled double-crystal monochromator: design and performances
Proux O., Nassif V., Prat A., Ulrich O., Lahera E., Biquard X., Menthonnex J.-J., Hazemann J.-L.
Journal of Synchrotron Radiation 13, 1 (2005) 59-68 - http://hal-insu.archives-ouvertes.fr/insu-00713235
Articles dans des revues avec comité de lecture
Physique/Physique/Instrumentations et Détecteurs
Feedback system of a liquid nitrogen cooled double-crystal monochromator: design and performances
Olivier Proux () 1, Vivian Nassif 2, Alain Prat 3, Olivier Ulrich 2, Eric Lahera 1, Xavier Biquard 2, Jean-Jacques Menthonnex 1, Jean-Louis Hazemann () 3
1 :  Observatoire des Sciences de l'Univers de Grenoble (OSUG)
http://osug.obs.ujf-grenoble.fr/
CNRS : UMS832 – INSU – Université Joseph Fourier - Grenoble I
414 Rue de la piscine - BP 53 38041 GRENOBLE CEDEX 9
France
2 :  Service de Physique des Matériaux et Microstructures (SP2M - UMR 9002)
http://inac.cea.fr/Phocea/Vie_des_labos/Ast/ast_service.php?id_unit=1
CEA : DSM/INAC – Université Joseph Fourier - Grenoble I – CNRS : UMR9002
SP2M - UMR 9002, Institut Nanosciences et Cryogénie, CEA/Grenoble, 17, Avenue des Martyrs 38042 Grenoble Cedex 9
France
3 :  Laboratoire de cristallographie (LC)
http://www-cristallo.grenoble.cnrs.fr/
CNRS : UPR5031
bat. F 25 Av des martyrs - BP 166 38042 GRENOBLE CEDEX 9
France
équipe FAME
SPMCE
Abstract We report on a new setup of indirect cryogenic cooling system for double-crystal monochromator running on BM30b / FAME beamline at the ESRF (Grenoble, France). This device has been conceived to limit the vibrations on the first diffracting crystal and to maintain it at a constant temperature. These points are crucial to maximize the beamline stability. Moreover, the relative angular position of the second crystal is dynamically adjusted by a piezoelectric transducer coupled with a feedback system in order to always be at the maximum of the rocking curve during a XAS scan. The temperature is stabilized with a 1/100° accuracy with two principal consequences. The energy resolution is close to the theoretical value (E/E=5.6 10-6 for Si(220)) and the precision of the energy positioning is extremely good even if the power load changes. A feedback mechanism allows a permanent and automatic optimisation of the angle between the two crystals of the monochromator. The intensity of the monochromatic beam remains optimized 1) when the intensity of the electron beam decreases in the storage ring and 2) during an energy scan.
Anglais
08/07/2005

Journal of Synchrotron Radiation (J Synchrotron Radiat)
Publisher International Union of Crystallography
ISSN 0909-0495 (eISSN : 1600-5775)
internationale
22/12/2005
13
1
59-68

double-crystal monochromator – cryogenic cooling – parallelism adjustment – X-ray absorption spectroscopy

Liste des fichiers attachés à ce document : 
PDF
2006_JSR_Proux.pdf(1.4 MB)